Evaluation of direct method for measuring low-density lipoprotein cholesterol
- VernacularTitle:低密度脂蛋白胆固醇直接测定法的评价
- Author:
Juan HE
;
Qinghua LI
;
Qiang DAI
- Publication Type:Journal Article
- From:
Chinese Journal of Clinical Laboratory Science
2001;19(3):152-154
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the precise,accurate and specific of two direct methods for measuring low-density lipoprotein cholesterol(LDL-C)based on the principle of selective hydrolysys. Methods Both DAIICHI and Randox reagents were compared with PVS method and the ultracentrifugally separated HDL and LDL fractions were used.Results Both methods all had good precise,the total CV was 3.96%~4.42%(Daiichi)and 0.78%~3.19%(Randox),repectively.The average concentrations of 48 serum samples were 3.68±1.23 mmol/L(PVS method),3.25±1.11 mmol/L(DAIICHI method)and 3.37±1.21 mmol/L(Randox method),respectively.There was no statistics difference between the results from PVS method and other two direct methods.Furthermore it indicated that the results determined by both direct methods had good relationship with that by PVS method.It also indicated that both direct methods had good specific to LDL-C.The dilute test demonstrated that there were good linearity between both direct methods of LDL-C and the linearity range was 9.28 mmol/L at least.Conclusion The direct methods for determining LDL-C based on the principle of selective hydrolysis possessed good precise and accurate and specific to LDL-C,and was meet with clinical application.