Flow Injection-Catalytic Spectrophotometric Determinationof Trace Phenols Using Stopped-Flow Technique
- VernacularTitle:废水中痕量酚的流动注射停流-催化光度法测定
- Author:
Yahong CHEN
;
Jing FAN
;
Cunling YE
;
Suling FENG
- Publication Type:Journal Article
- From:
Chinese Journal of Analytical Chemistry
2001;29(3):339-341
- CountryChina
- Language:Chinese
-
Abstract:
A rapid and simple flow-injection-catalytic spectrophotometric method using stopped-flow technique for determination of trace phenols is described. This methods is based on the catalytic effect of phenols on the reaction of rhodamine B and potassium bromate in sulfuric acid medium. The optimal experimental conditions are obtained by using the controlled and weighed centroid simplex method. The linear range of the determination is 0.60~4.00 mg/L for phenols, and the detection limit is 0.1 mg/L with the sample throughput of 34/h. The relative standard deviation of eleven replicate measurements is less than 2.0%. It has been applied for the determiantion of phenols in waste water with satisfactory results