The influence of different diluents on measurement of 15 items of biochemical parameters
10.3760/cma.j.issn.1673-4130.2009.03.003
- VernacularTitle:不同稀释介质对15个生化项目测定的影响
- Author:
Aiping ZHU
;
Xiuying LV
;
Shuyun GUO
- Publication Type:Journal Article
- Keywords:
Diluent;
Biochemical parameters;
Detection
- From:
International Journal of Laboratory Medicine
2009;30(3):214-216
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the influence of different diluents(physiologic saline, dis-tilled water and human inactivated serum) on measurement of 15 items of biochemical parameters.Methods Fifteen items of biochemical parameters [alanine transaminase(ALT), aspartic transaminase (AST), alkaline phosphatase( ALP), gamma glutamyl transpeptidase ( GGT), creatine kinase ( CK),lactate dehydrogenase(LD), hydroxybutyrate dehydrogenase(HBDH), total bilirubin(TBIL), direct bilirubin(DBIL), total bile acid(TBA), ereatinine(Cr), uric acid(UA), eholesterol(CHO), glucose (GLU), and blood urea nitrogen(BUN)] were chosen. For each parameter, 45 serum samples with different eoncentrations of the parameter were collected. After diluted with different diluents(physio-logic saline, distilled water or human inactivated serum), the serum samples were detected by applying the fully automated biochemical analyzer. The mean value was calculated and statistical analysis was performed. Results There were some differences of detection results when the specimens were diluted with different diluents. ALT, AST, GGT, DBIL, and HBDH serum samples could be diluted by 10 times with physiologic saline, distilled water or human inactivated serums ALP and TBA serum sam-ples could only be diluted with inactivated serum, otherwise its result would be lower; GLU, TBIL samples could be diluted with distilled water and inactivated serums for BUN, CR, UA, CK, LDH,and CHO samples, physiologic saline or human inactivated serum might be optimal; if distilled water was chosen, the results of other parameters tented to decline except UA. It was BUN was improper to dilute the BUN samples with distilled water. In addition, there was no significant difference between the items diluted by 5 times and 10 times with physiologic saline. All the 15 items could be diluted with inactivated serum. Conclusion The inactivated serum should be the first choice of diluents to e-nure the accurate results of biochemical parameters. If the prepared inactivated serum is absent, we may choose other diluents according to the above-mentioned results.