Application of Generalized Batho Method to Arbitrary Shape of Heterogeneous Tissues.
	    		
		   		
	    	
    	
    	
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Kyu Young CHAI
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
    Author Information Author Information
 
			        		
			        		
			        			1. Department of Therapeutic Radiology, College of Medicine, Seoul National University, Seoul, Korea.
 
 
- Publication Type:Original Article
- From:Journal of the Korean Society for Therapeutic Radiology
	            		
	            		 1987;5(2):165-168
	            	
            	
- CountryRepublic of Korea
- Language:Korean
- 
		        	Abstract:
			       	
			       		
				        
				        	The generalized Batho method, proposed by Webb and Fox, which is a method of calculation of dose correction factor for the multilayer of heterogeneous tissue, is complex even for a few kind of tissues. The method was modified for the purpose of getting a simple method that divide the multilayer of heterogeneous tissues into some groups of adjacent-tissue pairs. This new method could reduce the number of exponential terms and the time for calculating the dose correction factors by manual and computer calculation.