Determination of Glucose with Biosensor by Immobilization of Glucose Oxidase with Chitosan
- VernacularTitle:壳聚糖固定化葡萄糖氧化酶生物传感器测定葡萄糖的含量
- Author:
Yan ZHANG
;
Caifeng NAN
;
Li FENG
;
Liqin ZHANG
;
Chuan DONG
;
Shaomin SHUANG
- Publication Type:Journal Article
- Keywords:
Chitosan;
glucose oxidase;
glucose;
biosensor;
eggshell membrane
- From:
Chinese Journal of Analytical Chemistry
2009;37(7):1049-1052
- CountryChina
- Language:Chinese
-
Abstract:
A glucose biosensor was fabricated from a glucose oxidasE-immobilized by chitosan and oxygen electrode. The effects of concentration of chitosan(0.3%), enzyme loading(0.8 mg), pH 7.0, phosphate buffer concentration(300 mmol/L), and temperature 25 ℃ for the response of the biosensor were investigated. The glucose biosensor has a linear response range of 0.016-1.10 mmol/L with a detection limit of 8.0 μmol/L(S/N=3). The response time was less than 60 s. The biosensor showed extremely good stability with a shelf-life of at least 3 months. The biosensor exhibited good repeatable response to a 0.25 mmol/L glucose olution with a relative standard deviation of 2.5%(n=10). The precision of fabrication of the biosensors using four different membranes was good with a RSD of 4.7%. Some common potential components in sample such as niacinamide, vitamin B6, vitamin B12, vitamin E, Ca2+, Mg2+, K+ and Zn2+ showed no interferences on the response of the glucose biosensor. The biosensor was successfully applied to determine the glucose in commercial beverage samples.