Sequential Injection Instrument for Determination of Urine Iodine
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:顺序注射尿碘测定仪的研制及应用
- Author:
	        		
		        		
		        		
			        		Ying LIU
			        		
			        		;
		        		
		        		
		        		
			        		Rong ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Weijing BU
			        		
			        		;
		        		
		        		
		        		
			        		Jiebing TANG
			        		
			        		;
		        		
		        		
		        		
			        		Shan CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Lixia LIU
			        		
			        		
		        		
		        		
		        		
 
			        		
			        		
		        		 
- Publication Type:Journal Article
- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Urine iodine;
			        		
			        		
			        		
				        		sequential injection;
			        		
			        		
			        		
				        		catalysis kinetics;
			        		
			        		
			        		
				        		spectrophotometry
			        		
			        		
	        			
        			
        		
- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2009;37(11):1706-1710
	            	
            	
- CountryChina
- Language:Chinese
- 
		        	Abstract:
			       	
			       		
				        
				        	An iodine analysis instrument with sequential injection of urine samples was developed. A method for measurement of urine iodine was also developed by combining sequential injection and catalysis kinetics,and making use of catalysis facilitation of the iodine in the redox reaction of As~(3+) and Ce~(4+) . The sequential injection and stopped-flow stabilization determination were made possible by the program-controlled injector with controlled flow rate and the 16-hole program-controlled selection valves. The arsenic-cerium reaction with iodine-catalyzed at constant temperature state might proceed with the constant temperature flow cell. Using the syringes with program-controlled velocity,pushing and suction,Using the digital connected circuit and micro-iodine determination software,the reaction temperature of (32.0±0.1)℃,injection time of 45 s,stabilization time of 60 s,detection time of 20 s,injection volume of 400μL,linear range of 15 -600μg/L,detecting limitation of 5.01 μg/L(n=11) and recovery rate of 94. 1 % - 105. 1 % were obtained. With this method,the detecting result of the National Standard Reference (GBW09109 and GBW09110) materials was within a given standard range. Through this method,the detecting results had no significant differences comparing with those by standard method of National Health Service(P >0.05).