Feasibility and Stability of Right Ventricular Outflow Tract Pacing Under Current Technology
10.3969/j.issn.1000-3614.2009.03.012
- VernacularTitle:右心室流出道间隔部起搏的可行性和稳定性
- Author:
Ruohan CHEN
;
Keping CHEN
;
Fangzheng WANG
;
Wei HUA
;
Shu ZHANG
- Publication Type:Journal Article
- Keywords:
Right ventricular outflow tract;
Active fixation;
Pacing4
- From:
Chinese Circulation Journal
2009;24(3):202-205
- CountryChina
- Language:Chinese
-
Abstract:
Objectives: To assess the feasibility and stability of right ventricular outflow tract (ROVT) pacing under current technology by comparing the results of ROVT pacing with the traditional right ventricular apex (RVA) pacing. Methods: A total of 42 patients (at mean age of 63.5±10.4 years) without structural heart disease were randomly divided into two groups. RVA pacing group (n=14),and RVOT pacing group(n=28). An active fixation lead was implanted in all patients whose pacemaker could automatically measure the pacing threshold every day. The operation time,X-ray exposure time and lead parameters detected during the operation were collected to evaluate the feasibility of RVOT pacing. The complications related to lead and implantation procedure and the trend of threshold change during the follow-up time were used to assess the stability of RVOT pacing.Results: There were no statistic differences between RVA pacing group and RVOT pacing group in terms of operation time,X-ray exposure time and lead parameters. In RVOT group,the change of threshold during acute period was similar to those in RVA group (P=0.23). Chronic pacing threshold was also comparable between two groups,mean threshold at 6 months follow-up time was 0.55±0.11V and 0.54±0.09V at 0.4 pulse width in RVA group and RVOT group respectively (P=0.787).Conclusion: RVOT pacing was feasible and stable in operation time and lead characteristics compared with the conventional RVA pacing under current pacing technology.