Electromagnetic navigation-guided acupuncture of foramen ovale: A cadaver study
10.3969/j.issn.1001-3733.2010.01.012
- VernacularTitle:电磁导航卵圆孔穿刺术的实验研究
- Author:
Minjie CHEN
;
Chi YANG
;
Lixu GU
;
Weijie ZHANG
;
Minjun DONG
;
Jing ZHAO
;
Bin JIANG
;
Yating QIU
;
Zhiqiang FENG
- Publication Type:Journal Article
- Keywords:
Electromagnetic navigation;
Free-hand puncture;
Foramen ovale;
Cadaver study
- From:
Journal of Practical Stomatology
2010;26(1):55-59
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To evaluate the surgView-RFT electromagnetic navigation system in radiofrequency thermocoagulation(RFT). Methods: 6 foramen ovales of 3 cadavers were punctured by 3 beginners. Every one tried 5 times by free-hand and 5 times by SurgView-RFT electromagnetic navigation-guidance. After each puncture, CT scans were used as golden standard to measure the distance between the tip of needle and the anterior margin of the foramen ovale. T test and variance of SAS 6.12 statistical software was used to do the statistic analysis. Results: The punctures by free-hand failed, and the distances between the tip of needle and the anterior margin of the foramen ovale were 8-10.6 mm. The punctures by navigation guidance were successful, and the distances between the tip of needle and the anterior margin of the foramen ovale were 2.68-3.54 mm. The error of navigation system was 0.59 mm which was verified by CT scan. There was significant difference between free-hand puncture and navigation-guided puncture. Conclusion: The surgView-RFT electromagnetic navigation system is characteristic with high performance, high accuracy and minimal invasion, while it is applied in training or clinical using.