Evaluate the applicability of the international criteria for blood smear review on the hematology analyzer
10.3760/cma.j.issn.1673-4130.2010.05.007
- VernacularTitle:血涂片复审准则在血细胞分析仪中的适用性评估
- Author:
Xiaoping LI
;
Qikang WU
;
Zhiqi QIU
- Publication Type:Journal Article
- Keywords:
blood cell count;
smear layer;
Equipment and Supplies
- From:
International Journal of Laboratory Medicine
2010;31(5):432-433
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the applicability of the international criteria for blood smear review on the SysmexXE-2100 hematology analyzer.Methods 300 blood samples were tested by hematology analyzer Sysmex-XE2100,then it were divided into two groups,that is,sieving group and sieved group according to the criteria of international boold smear review,then blood smears were reviewed with optical microscope.Results Among 300 samples,,54 cases were true positive(18%),true negative was 64%,48 cases were false positive(16%),four 6cases were false-negative (2%),the accuracy rate was 82%.Conclusion Performance of microscopic observation after analyzer detection according to the international criteria for blood smear review presented good efficiency,it will show better practical value analysising with the clinical condition.