Three-dimensional dose rate table for 103pd radioactive stent
- VernacularTitle:103Pd放射性支架的三维剂量率分布
- Author:
Zhiyong XU
;
Liangan ZHANG
;
Guangfu DAI
;
Shuyu FAN
- Publication Type:Journal Article
- Keywords:
Radioactive stent;
Restenosis;
Dose kernel function;
Monte Carlo code
- From:
Chinese Journal of Radiation Oncology
2008;17(3):230-233
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate three-dimensional dose distribution for 103Pd radioactive stent.Methods The surface dose,the axial dose and radial dose in surface for 103pd stent (3 mm × 13 mm) were estimated by experimental simulating method, analytic function and MCNP4b code. Three-dimensional dose distribution was calculated by MCNP4b code. Results The surface dose of 103pd stent was 0. 109 and 0. 106 Gy estimated by experimental simulating method and MCNP4b code,between which the difference was less than 3%. The axial dose calculated by analytic function and MCNP4b code was well consistent,and so was the radial dose estimated by the three methods. Dose rate table were estimated by MCNP4b code. Conclnsions Dose distribution for 103 Pd stent estimated by the three methods is relatively accurate. Three-dimensional dose table estimated by MCNP4b may be used to calculate dose for 103Pd stent in animal experiment and clinical application.