Comparative study between right ventricular septal pacing and right ventricular apical pacing
- VernacularTitle:右室间隔部主动电极和右室心尖部被动电极的对照研究
- Author:
Huijun LI
;
Haibo QU
;
Zhanquan LI
- Publication Type:Journal Article
- Keywords:
Heart ventricle;
Heart septum;
Electrodes,implanted;
Cardiac apex;
Pacing
- From:
Chinese Journal of Postgraduates of Medicine
2008;31(16):11-13
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the feasibility and safety of right ventricular septal pacing with active fixation electrodes.Methods This was a randomized and control clinical trial.Patients implanted with pacemaker were randomly divided into two groups.One group underwent the right ventrieular high septal pacing with the active fixation electrodes(RVS group),and the other group underwent the right ventricular apical pacing with the passive fixation electrodes(control group).The parameter was recorded in two groups in and after operation and compared accordingly.Results (1)There were no significant differences in the procedure time between two groups[averaged(59.6±3.2)and(60.2±3.7)minutes respectively].But the time of exposing X-ray was significant longer in RVS group.(2)There was no significant difference in acute implanting measurement,except perioperative and postoperative threshold 1 month in RVS group was higher than that in control group(P<0.05 or<0.01).After 3 months,there was no significant difference.(3)Impedance decreased significantly in 1 month and 3 months in RVS group[P<0.01 or<0.05).(4)There was no difference in R-wave sensing between two groups.(5)After 100%pacing,the mean QRS duration Was shorter in RVS group,but the difference was not significant.(6)Operations in two groups came off smoothly and there were no complications.Conclusion It is feasible and secure to pace from RVS with active fixation electrodes.