Treatment verification of helical tomotherapy intensity modulated radiation therapy
- VernacularTitle:螺旋断层放疗系统调强放疗验证
- Author:
Shouping XU
;
Xiaowu DENG
;
Xiangkun DAI
;
Lianyuan WANG
;
Chuanbin XIE
;
Ruigang GE
;
Xiangyan SHA
- Publication Type:Journal Article
- Keywords:
Tomotherapy;
Intensity modulated radiotherapy;
Dose verification;
Quality as-Surance
- From:
Chinese Journal of Radiation Oncology
2008;17(5):395-397
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the clinical feasibility of quality assurance of helical tomotherapy intensity modulated radiation therapy (IMRT) through a series of clinical case studies. Methods Tomotherapy planning system was used to provide optimized IMRT treatment plans. To verify the dose of IMRT plans,the cylindrical Virtual Water phantom,0.056cm3 AISL, ion chamber and EDR2 film were used for getting the absolute and relative doses from tomotherapy IMRT planning. The film and ion chamber were placed into the phantom. The doses of the interesting points and isodose distribution of the axial plane were measured,then the results were compared with those from the calculation in planning system for verification. Before the irradiation,kVCT images of the phantom were registered by tomotherapy MVCT images. So the setup of phantom was guided for verifying the position. Results The isedose distribution (on the axial plane) measured by the film was well consistent with that calculated by tomotherapy planning system. The error between the measured dose and predicted dose in the measured points was less than 3%. The setup error of the phantom was able to be kept within 1 mm. There was 2 mm offset along the vertical direction of couch from the virtual isocenter to beam isocenter,which should be considered during the phantom/patient setup. Conclusions The procedures for quality assurance of tomotherapy IMRT are feasible in our experience. And helical tomotherapy IMRT QA system has been constituted.