Intracranial aneurysms on dual-source CT angiography:comparison with conventional and three dimensional DSA
- VernacularTitle:颅内动脉瘤双源CT血管成像与常规和三维DSA结果的比较
- Author:
Yane ZHAO
;
Longjiang ZHANG
;
Changsheng ZHOU
;
Wei HUANG
;
Guangming LU
- Publication Type:Journal Article
- Keywords:
Intracranial aneurysm;
Tomography,X-ray computed;
Angiography,digital subtraction
- From:
Chinese Journal of Radiology
2008;42(9):941-944
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the value of dual-source computed tomography angiography (DSCTA)in detecting intracranial aneurysms by comparing with conventional and three-dimensional DSA.Methods In this study,95 patients with subarachnoid hemorrhage(SAH)underwent both DSCTA and DSA examination.The detection rate,size,and ratio of the neck to the dome(N/D ratio)of the aneurysrns were evaluated.Statistical analysis was performed using a paired sample Student's t-test for the comparisons of the value of N/D and 2 Related Samples test for long axis.Results A total of 67 aneurysms in 63 patients at DSA and 64 aneurysms in 60 patients at DSCTA were detected,respectively;whereas no aneurysm was detected in 32 patients at DSA.Compared with DSA,the overall sensitivity.specificity,positive predictive value,and negative predictive value of DSCTA on a per-aneurysm basis were 94.2%,100.0%,100.0%,and 91.4%,respectively.For the aneurysms larger than 3 mm,the sensitivity and specificity of DSCTA in detecting intracranial aneurysms were equal to those of DSA:For aneurysms smaller than 3 mm,however,the sensitivity and specificity of DSCTA is 80.0% and 100.0%.The N/D ratio for DSA and DSC:TA was 0.46±0.14 and 0.51±0.18.respectively,and the median of long axis was 4.9 mm and 4.8 mm.respectively.There was no significant difierence in the N/D ratio(t=3.20;P>0.05)and the long axis(Z=-1.309;P>0.05)between DSA and DSCT.Condusions Compared with conventional and three-dimensional digital subtraction angiography,DSCTA has high sensitivity and specificity in the detection of intracranial aneurysms,especially for detection of snlall aneurysnm(<3 mm).It can be used as a routine screening technique.