Comparative study of image quality and radiation dose between prospective and retrospective ECG gating technique in coronary artery imaging with 64-slice spiral CT
10.3760/cma.j.issn.0254-5098.2009.03.028
- VernacularTitle:64排螺旋CT冠状动脉成像的前瞻性与回顾性心电门控技术的图像质量及辐射剂量的比较研究
- Author:
Jianxin LIU
;
Jian LIU
;
Yanbin DOU
;
Jichen WANG
;
Hongyue SUN
- Publication Type:Journal Article
- Keywords:
64-slice spiral CT;
ECG gating;
Coronary artery;
Radiation dose
- From:
Chinese Journal of Radiological Medicine and Protection
2009;29(3):320-323
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the image quality and radiation dose between prospective ECG-trigering and retrospective ECG gating technique in coronary artery imaging. Methods 33 patients suspected coronary artery disease were included in this study and divided into experimental group (prospective ECG-triggering coronary artery imaging, heart rate < 65 bpm, 16 patients)and control group (retrospective ECG gating coronary artery imaging, heart rate < 75 bpm, 17 patients). Radiation dose was recorded for statistical analysis. At the same time, image quality was evaluated. Results There were no significant differences in image quality 3.2 msy. Effective dose reduced 76.47 %. Conclusions Prospective ECG gating coronary artery imaging can obtain the similar image quality compared with prospective ECG gating coronary artery imaging but the effective dose reduced 76.47 %. Prospective ECG gating coronary artery imaging has clinical value to peoples who are able to not accept high radiation dose and with low heart rate.