Application of FLOW-PRA detection in clinical organ transplantation
10.3760/cma.j.issn.0254-1785.2010.03.011
- VernacularTitle:流式细胞术-群体反应性抗体的检测在临床器官移植中的应用
- Author:
Wei HUANG
;
Zhishui CHEN
;
Weijie ZHANG
;
Dunxiu CHENG
;
Xing WANG
;
Dawei WANG
;
Ping ZHOU
- Publication Type:Journal Article
- Keywords:
Flow cytometry;
Enzyme-linked immunosorbent assay;
PRA
- From:
Chinese Journal of Organ Transplantation
2010;31(3):173-176
- CountryChina
- Language:Chinese
-
Abstract:
Objective To apply flow cytometry-panel reactive antibody (FLOW-PRA) and compare the application of traditional (enzyme-linked imrnunosorbent assay) ELISA-PRA in clinical organ transplantarion,so as to evaluate the concordance,sensitivity,accuracy and practicability of FLOW-PRA.Methods PRA was detected in 212 serum samples from 185 patients awaiting organ transplantation using FLOW-PRA and ELISA-PRA.Results It took 1.5 h and 3 h for FLOW-PRA vs ELISA-PRA.Concordance correlation coefficient for the results of the two methods was 94% (class Ⅰ)and 89% (class Ⅱ),respectively.Of all sera,24.5% (in comparison to ELISA-PRA,P<0.005)were class I positive,18.4% (P<0.05) class Ⅱ positive by flow cytometry,and 17.9% and 14.6% by ELISA,respectively.The positive incidence in Flow group was higher than in ELISA group.Low titer of antibodies was detected positively only by flow eytometry,furthermore,the antigen specificity of PRA could only be discriminated by FLOW-PRA.Conclusion Flow cytometry is more sensitive and more accurate than ELISA in PRA detection.FLOW-PRA is easy to operate and time-effective,and suitable for clinical application.