Impact of thermoplastic mask on X-ray surface dose calculated with Monte Carlo code
10.3760/cma.j.issn.1004-4221.2010.04.018
- VernacularTitle:热塑膜对X射线治疗剂量影响的研究
- Author:
Yanqun ZHAO
;
Jie LI
;
Liping WU
;
Pei WANG
;
Jinyi LANG
;
Dake WU
;
Mingyong XIAO
- Publication Type:Journal Article
- Keywords:
X-rays;
Thermoplastic mask;
Dosimetry;
Monte Carlo code
- From:
Chinese Journal of Radiation Oncology
2010;19(4):336-339
- CountryChina
- Language:Chinese
-
Abstract:
Objective To calculate the effects of thermoplastic mask on X-ray surface dose.Methods The BEAMnrc Monte Carlo Code system, designed especially for computer simulation of radioactive sources, was performed to evaluate the effects of thermoplastic mask on X-ray surface dose.Thermoplastic mask came from our center with a material density of 1.12 g/cm2. The masks without holes,with holes size of 0. 1 cm× 0. 1 cm, and with holes size of 0. 1 cm × 0. 2 cm, and masks with different depth (0.12 cm and 0.24 cm) were evaluated separately. For those with holes, the material width between adjacent holes was 0. 1 cm. Virtual masks with a material density of 1.38 g/cm3 without holes with two different depths were also evaluated. Results Thermoplastic mask affected X-rays surface dose. When using a thermoplastic mask with the depth of 0. 24 cm without holes, the surface dose was 74. 9% and 57.0% for those with the density of 1.38 g/cm3 and 1.12 g/cm3 respectively. When focusing on the masks with the density of 1.12 g/cm3, the surface dose was 41.2% for those with 0.12 cm depth without holes;57.0% for those with 0. 24 cm depth without holes;44. 5% for those with 0. 24 cm depth with holes size of 0.1 cm ×0.2 cm;and 54.1% for those with 0.24 cm depths with holes size of 0.1 cm ×0.1 cm.Conclusions Using thermoplastic mask during the radiation increases patient surface dose. The severity is relative to the hole size and the depth of thermoplastic mask. The surface dose change should be considered in radiation planning to avoid severe skin reaction.