Quantification of fat using chemical shift imaging and 1H-MR spectroscopy in phantom model
10.3760/cma.j.issn.1005-1201.2010.07.016
- VernacularTitle:化学位移成像及MR波谱分析在体外水脂模型中对脂肪定量的研究
- Author:
Xingui PENG
;
Shenghong JU
;
Fang FANG
;
Gaojun TENG
- Publication Type:Journal Article
- Keywords:
Magnetic resonance imaging;
Magnetic resonance spectroscopy;
Fats;
Models
- From:
Chinese Journal of Radiology
2010;44(7):747-752
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the accuracy of chemical shift imaging (CSI) and MR spectroscopy (MRS) for fat quantification in phantom model. Methods Eleven phantoms were made according to the volume percentage of fat ranging from 0 to 100% with an interval of 10% . The fat concentration in the phantoms were measured respectively by CSI and MRS and compared using one-sample t test The correlation between the two methods was also analyzed. The concentration of saturated fatty acids ( FS), unsaturated fatty acids (FU) and the polyunsaturation degree (PUD) were calculated by using MRS. Results The fat concentration was (48.0 ±1.0)%, (57.0 ±0.5)%, (67.3±0.6)%, (77.3 ±0.6)%, (83.3±0.6)% and (91.0±1.0)% respectively with fat volume of 50% to 100% by CSI. The fat concentration was (8.3 ±0.6)%, (16.3 ±0.7)%, (27.7 ±0.6)%, (36.0 ±1.0)%, (43.5±0. 6)% and (56. 5 ±1. 0)% respectively with fat volume of 10% to 60% by MRS, the fat concentration were underestimated by CSI and MRS (P < 0.05 ), and had high linear correlation with the real concentration in phantoms ( CSI: r = 0. 998, MRS: r = 0.996, P < 0.01 ) . There was also a linear correlation between two methods (r = 0. 992, P < 0. 01 ) but no statistically significant difference ( pairedsamples t test, t = -0. 125 ,P = 0.903). By using MRS, the relative ratio of FS and FU in fat were 0. 15and 0. 85, the PUD was 0. 0325, respectively, and highly consistent with these in phantoms. Conclusion Both CSI and MRS are efficient and accurate methods in fat quantification at 7.0 T MR.