Application value of computed radiography energy subtraction in detecting pulmonary soft-tissue:comparison with direct digital radiography
10.3760/cma.j.issn.0254-5098.2010.04.032
- VernacularTitle:计算机X线摄影能量减影与直接数字摄影对胸部软组织病变检测能力比较
- Author:
Menglong ZHANG
;
Kai LIU
;
Liangan ZHANG
- Publication Type:Journal Article
- Keywords:
Energy subtraction;
Computed radiography ( CR );
Direct digital radiography (DR);
Comparative study
- From:
Chinese Journal of Radiological Medicine and Protection
2010;30(4):483-485
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the efficiency of computed radiography energy subtraction (ES-CR) with direct digital radiography (DR) in detecting the pulmonary soft-tissue signals.Methods The adult chest phantom including TRG was imaged by ES-CR and DR separately and was estimated by 6 radiologists with 5-value-differentiation method. ROC was used to analyze the performance of two systems. Images of DR and ES-CR of 28 patients with small lung cancer were diagnosed by 3 radiologists. Results The area under the ROC curves was 0. 727 in ES-CR and 0.827 in DR (Z= 2.96,P < 0. 05). 28 small lung cancer patients were significantly different between two systems (x2 = 5.14,P <0. 05). Conclusions The true positive proportion of DR might be higher than that of ES-CR. ES-CR is inferior to DR in detecting pulmonary soft-tissue signals.