Preliminary study of MR imaging in rat with series coil
10.3760/cma.j.issn.1005-1201.2010.09.024
- VernacularTitle:串联线圈大鼠MR成像的初步应用研究
- Author:
Xiubin WANG
;
Hetao CAO
;
Min LI
;
Hongmei JING
;
Yongjin PAN
- Publication Type:Journal Article
- Keywords:
Magnetic resonance imaging;
Magnetic resonance spectroscopy;
Spinal cord injury;
Animal experimentation
- From:
Chinese Journal of Radiology
2010;44(9):991-994
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the effect of MR imaging in rats with series coil. Methods MR scanning was performed on the self-made aqueous solution phantom with the series coil and other three coils respectively, using the same FRFSE- XL pulse sequence. SNR of the same image was obtained in the same position slice for all four coils. Twelve SD rats were divided into three model groups and one normal group. FRFSE-XL FSPGR and Probe-SI pulse sequence were respectively used to show the brain injury,1H-MRS, spinal cord injury (SCI) and abdomen with series coils. Results The SNR of series coil (39. 7)was over six times higher than the SNR of knee joint coil (6. 41 ), which was the best in other three coils.Grey matter and white matte; the brain structure and hematomas were clearly depicted in the rats'brain on T1WI and T2WI. From the second week after injection of 6-OHDA into the rats brain, the ratio of NAA/Cr decreased (before injection 1.24 ,two weeks after injection 0. 781 ). The spinal cord and the configuration of the white matter were clearly showed on T2WI image, and the location and degree of SCI in the rat thoracic spinal cord were also well defined on T2 WI image. The pulse sequence of FSPGR T1 WI was used for rat abdomen scanning. The scan time was only eight seconds. It is helpful to overcome the breathing movement effect and show the structure of rat abdomen. Conclusion It provided a good means for the rat MR in vivo to make use of 3.0T MR combined with series coils.