Quality control and quality assurance of single photon emission computed tomography/computed tomography ( SPECT/CT) system
10.3760/cma.j.issn.1008-6706.2010.19.004
- VernacularTitle:单光子发射计算机体层摄影术/计算机断层扫描的质量控制与质量保证
- Author:
Yihua LIANG
;
Weiwu QIN
;
Jian YE
;
Dangsheng LI
;
Zhangwei LUO
;
Liyan DENG
;
Miao HUANG
- Publication Type:Journal Article
- Keywords:
Tomography;
Emission computed,Single photon;
Quality control
- From:
Chinese Journal of Primary Medicine and Pharmacy
2010;17(19):2600-2602
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the quality control and quality assurance of the SPECT/CT system.Methods The energy peak,energy resolution capacity and the intrinsic uniformity of the device's detector were regularly examined, and the quality control was performed. Results The daily hardware maintain could reduce the rate of system's trouble. The device's energy peak,energy resolution capacity were consistent during half year's observation period. The two detector's intrinsic uniformity were better after calibration than before. [ detector Ⅰ: ( 2.71 ± 0.28 ) vs (2.37±0.11)(t=2.489,P<0. 05);detector Ⅱ:(2.68 ±0.12)vs(2.38 ±0. 19)(t =6.421,P <0.01) ] .Conclusion Regular quality control and maintain could keep the function stabilization,enhance the availability rate of the SPECT/CT system.