Energy dependence of three passive solid detectors on low energy X-ray
10.3760/cma.j.issn.0254-5098.2010.06.028
- VernacularTitle:3种无源固体探测器对低能X射线能量响应的研究
- Author:
Jianchao WANG
;
Hui XU
;
Kedao WEI
;
Baorong YUE
- Publication Type:Journal Article
- Keywords:
Mammography;
Solid detector;
Energy dependence;
Dose measurement
- From:
Chinese Journal of Radiological Medicine and Protection
2010;30(6):735-736
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray.Methods Two thermoluminescent TL) detectors ( LiF(Mg,Ti) and LiF( Mg,Cu, P) and one optically stimulated luminescence (OSL) detector( Al2O3: C ) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography.