The analytical performance assessment of high-sensitivity cardiac troponin T reference intervals investigation
10.3760/cma.j.issn.1009-9158.2010.09.003
- VernacularTitle:高敏感方法检测心肌肌钙蛋白T的分析性能评价
- Author:
Lingyan SONG
;
Jiong WU
;
Binbin SONG
;
Wenqi SHAO
;
Chunyan ZHANG
;
Beili WANG
;
Wei GUO
;
Baishen PAN
- Publication Type:Journal Article
- Keywords:
Troponin T;
Reference values;
Diagnostic techniques and procedure;
Evaluation studies
- From:
Chinese Journal of Laboratory Medicine
2010;33(9):814-818
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the analytical performance of hs-cTnT and biological variations in healthy population as well as establish hs-cTnT reference intervals. Methods The serum samples from 100 acute myocardial infraction patients and 474 apparently healthy subjects were collected. The functional sensitivity,within- and between-run imprecision were determined. The hs-cTnT assay and con-cTnT assay were evaluated. The serum hs-cTnT levels were detected in apparently healthy subjects to establish reference intervals. Moreover,the long-term and short-term biological variations for hs-cTnT in healthy volunteers were assessed. Results The functional sensitivity of hs-cTnT was 0. 005 μg/L. The within- and between-run precision for lower level control(0. 014 μg/L) and higher level control(2. 500 μg/L) was 2. 97% vs 3. 64%and 0. 66% vs 1.01% ,respectively. The correlation between hs-cTnT assay and con-cTnT assay was good ( R2 =0. 972 ,P <0. 01 ). The 99th percentile in apparently healthy subjects was 0. 003 μg/L for women less than60 years, 0.008 μg/L for men less than 60 years, 0.015 μg/L for women above 60 years and 0. 021 μg/L for men above 60 years. The CVa, CVi, CVg and CVt of short-term biological variations in detecting hs-cTnT from 22 apparently healthy subjects were 3.8%, 4. 8%, 49.9% and 58.5%,respectively. The CVa,CVi ,CVg and CVt of long-term biological variations were 5. 3% ,6. 4% ,56. 6% and 68. 3% respectively. Conclusions The analytical performance of the hs-cTnT is better than con-cTnT assay,achieving acceptable level according to guideline. Our experimental result could provide the basis for the new high sensitivity cTnT assay in the diagnosis of acute myocardial infarction.