Frequency of endoscopic ultrasonography for diagnosis of duodenal submucosal lesions
10.3760/cma.j.issn.1007-5232.2010.02.005
- VernacularTitle:十二指肠隆起性病变超声内镜频率选择的临床研究
- Author:
Hong ZHU
;
Ruihua SHI
;
Xiayue HUANG
;
Lianzhen YU
;
Xueliang LI
;
Guoxin ZHANG
;
Shuping YANG
;
Qiyun TANG
;
Jing DING
- Publication Type:Journal Article
- Keywords:
Gastric mucosa;
Duodenum;
Endoscopic ultrasonography;
Diagnosis
- From:
Chinese Journal of Digestive Endoscopy
2010;27(2):71-74
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the effect of different frequencies of endoscopic ultrasonography (EUS) on diagnosis of duodenal submucesal lesions. Methods A total of 210 elevated duodenal lesions detected by routine endoscopy in 201 patients were examined by EUS at different frequencies according to lesion features and image clarity. Endoscopic resection was applied in 27 lesions with informed consents,and pathological findings were compared with the results of EUS. Results The accuracy of EUS was 81.4% (22/27). Ultrasound of low frequency was superior in displaying overview of lesion and was more suitable for lesions larger than 1.5 cm in diameter or lesions of hyperecho. On the other hand,ultrasound of high frequency was more reliable to reveal minute structures and could be employed in scanning lesions smaller than 1.5 cm in diameter. Combination of different frequencies of ultrasound could provide more details in larger lesions. Conclusion Application of appropriate frequency or combined frequencies of ultrasound in examination of elevated duodenal lesions is beneficial to achieve more accurate diagnosis.