A Case of Acute Aortic Regurgitation due to Leaflet Dehiscence of the Carpentier-Edwards Pericardial Bioprosthesis 18 Years after Implantation
10.4326/jjcvs.43.344
- VernacularTitle:大動脈弁位生体弁置換術後18年目に人工弁断裂に起因する急性心不全を来した1例
- Author:
Satoru Otani
;
Tsuyoshi Yamamoto
;
Yuki Yamada
;
Taiichiro Matsumoto
- Publication Type:Journal Article
- Keywords:
acute aortic regurgitation;
pericardial bioprosthesis;
bovine pericardial valve;
leaflet dehiscence
- From:Japanese Journal of Cardiovascular Surgery
2014;43(6):344-346
- CountryJapan
- Language:Japanese
-
Abstract:
A 65-year-old man, who had undergone the aortic valve replacement with a Carpentier-Edwards pericardial bioprosthesis (CEP 25 mm) 18 years previously (at age 48), was admitted to our hospital with a diagnosis of acute heart failure due to acute aortic regurgitation. Redo surgery was performed. The ascending aorta was cross clamped, and cardiac arrest was induced, and aortotomy was done. One of the leaflets of the CEP was entirely collapsed and dislocated to the LV side, which caused acute aortic regurgitation. Although there was no evidence of endocarditis, the other two leaflets of CEP were severely calcified. Aortic valve replacement was performed with a CEP 23 mm. He was discharged in good condition on the 16th post-operative day.