Bioprosthetic Valve Dysfunction due to Pannus after Tricuspid Valve Replacement
10.4326/jjcvs.43.97
- VernacularTitle:パンヌス形成による三尖弁位生体弁機能不全に対し再弁置換術を施行した1例
- Author:
Seiji Matsukuma
;
Kiyoyuki Eishi
;
Koji Hashizume
;
Tsuneo Ariyoshi
;
Shinichiro Taniguchi
;
Mizuki Sumi
- Publication Type:Journal Article
- Keywords:
tricuspid valve replacement;
prosthetic valve failure;
pannus
- From:Japanese Journal of Cardiovascular Surgery
2014;43(3):97-100
- CountryJapan
- Language:Japanese
-
Abstract:
Prosthetic valve dysfunction due to pannus formation is an infrequent but serious complication of tricuspid valve replacement. An 87-year-old woman underwent tricuspid valve re-replacement for severe prosthetic valve stenosis and regurgitation. On removal, thick fibrous pannus and chordal attachments were observed on the ventricular side of the cusp, which corresponded to the septal leaflet of the native valve. Microscopic examination revealed inflammatory cell infiltration accompanied with severe fibrosis and scarring had compromised and broken the prosthetic valve cusp under the pannus. The elastic fiber, which was detected in the base of the pannus, suggested it was a remnant of the native tricuspid valve leaflet. Prevention of native tissue attachment to the prosthetic valve cusp, which may cause severe pannus formation, appears to be extremely important for the long-term outcome and valve durability. The choice of prosthesis for the tricuspid position remains controversial. We should especially consider the height of stent posts and the continuity between the cusp and suture ring in the choice of the bioprosthetic valve for tricuspid position.