Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens.
10.3343/alm.2015.35.3.341
- Author:
Gilsung YOO
1
;
Juwon KIM
;
Young UH
;
Kwang Ro YOON
;
Soon Deok PARK
;
Kap Jun YOON
Author Information
1. Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Korea. u931018@yonsei.ac.kr
- Publication Type:Original Article ; Research Support, Non-U.S. Gov't
- Keywords:
Hemolysis;
Score;
System;
Anticoagulants;
Analyzer
- MeSH:
Anticoagulants/*pharmacology;
*Blood Specimen Collection;
Edetic Acid/pharmacology;
Hemoglobins/analysis;
Hemolysis/drug effects;
Humans
- From:Annals of Laboratory Medicine
2015;35(3):341-347
- CountryRepublic of Korea
- Language:English
-
Abstract:
BACKGROUND: The identification of in vitro hemolysis (IVH) using a hematology analyzer is challenging because centrifugation of the specimens cannot be performed for cell counts. In the present study, we aimed to develop a scoring system to help identify the presence of hemolysis in anticoagulated blood specimens. METHODS: Thirty-seven potassium EDTA anticoagulated blood specimens were obtained, and each specimen was divided into 3 aliquots (A, B, and C). Aliquots B and C were mechanically hemolyzed by aspirating 2 and 5 times, respectively, using a 27-gauge needle and then tested; aliquot A was analyzed immediately without any hemolysis. After the cells were counted, aliquots B and C were centrifuged and the supernatants were tested for the hemolytic index and lactate dehydrogenase levels. RESULTS: The 4 hematologic parameters were selected and scored from 0 to 3 as follows:< 34.0, 34.0-36.2, 36.3-38.4, and > or =38.5 for mean cell hemoglobin concentration (MCHC, g/dL); <0.02, 0.02, 0.03, and > or =0.04 for red blood cell ghosts (10(12)/L); <0.13, 0.13-0.38, 0.39-1.30, and > or =1.31 for difference value (g/dL) of measured hemoglobin and calculated hemoglobin; and <0.26, 0.26-0.95, 0.96-3.34, and > or =3.35 for difference value (g/dL) of MCHC and cell hemoglobin concentration mean. The hemolysis score was calculated by adding all the scores from the 4 parameters. At the cutoff hemolysis score of 3, the IVH of aliquots B and C were detected as 64.9% and 91.9%, respectively. CONCLUSIONS: The scoring system might provide effective screening for detecting spurious IVH.