Perigraft Seroma after Endovascular Repair of the Abdominal Aortic Aneurysm
10.4326/jjcvs.32.285
- VernacularTitle:腹部大動脈りゅうステントグラフト手術後のperigraft seroma
- Author:
Naoki Toya
;
Takao Ohki
;
Makoto Sumi
;
Hisano Toriumi
;
Yuka Negishi
- Publication Type:Journal Article
- Keywords:
seroma
- From:Japanese Journal of Cardiovascular Surgery
2003;32(5):285-287
- CountryJapan
- Language:Japanese
-
Abstract:
A 75-year-old man underwent endovascular repair of the abdominal aortic aneurysm. The procedure was aorto-uni-femoral endograft and femorofemoral crossover bypass using PTFE graft. The postoperative course was satisfactory, but 4 years after operation, he was admitted complaining of abdominal fullness. CT scan showed significant increase of aneurysm diameter to 13cm without evidence of endoleak. Endograft replacement with a new Dacron graft was carried out. Intraoperative findings suggested perigraft seroma related to the use of PTFE, and there was no endoleak. The postoperative course was uneventful. Pathological finding of aneurysm showed a lack of hemocytes and thrombocytes.