Dysfunction of Bioprosthetic Valve Presenting with Musical Cooing Murmur. Report of a Case.
10.4326/jjcvs.20.857
- VernacularTitle:Cooing murmurを伴う僧帽弁位生体弁機能不全の1手術治験例
- Author:
Takeshi MIYAIRI
;
Yutaka KOTSUKA
;
Ryushi MURAKAMI
;
Jun NAKAJIMA
;
Hideto NAKAHARA
;
Akira MIZUNO
- Publication Type:Journal Article
- From:Japanese Journal of Cardiovascular Surgery
1991;20(5):857-860
- CountryJapan
- Language:Japanese
-
Abstract:
A case of xenograft valve dysfunction presenting with musical cooing murmur is reported. The patient was a 47-year-old woman and had received mitral valve replacement with porcine xenograft (Carpentier-Edwards 31-M) 10 years before Preoperative echocardiogram showed vibration of a leaflet outside the stent of the bioprosthetic valve. Cardiac catheterization revealed stenosis as well as regurgitation of the bioprosthetic valve. The resected xenograft valve showed a tear which produced regurgitation and musical cooing murmur in one of three pliable leaflets and it also showed pannus formation which caused stenosis.