Early Experience with the 19-mm Medtronic Mosaic Porcine Bioprosthesis for Small Aortic Annuli
10.4326/jjcvs.37.1
- VernacularTitle:狭小大動脈弁位における19mm Mosaic 弁の短期成績について
- Author:
Hiroshi Kagawa
;
Yoshimasa Sakamoto
;
Hiroshi Okuyama
;
Shinichi Ishii
;
Shingo Taguchi
;
Kazuhiro Hashimoto
- Publication Type:Journal Article
- From:Japanese Journal of Cardiovascular Surgery
2008;37(1):1-5
- CountryJapan
- Language:Japanese
-
Abstract:
A study was conducted to evaluate the clinical and hemodynamic performance of the 19-mm Medtronic Mosaic Valve (MMV) in the aortic position, which is a third-generation stented porcine bioprosthesis. Between 2003 and 2006, 9 patients underwent AVR using the 19-mm MMV. None of the patients were suitable for a 19-mm Perimount bioprosthetic valve due to having a small annulus and sinotubular junction. The patients included 3 men and 6 women with a mean age of 73.2±4.97 years and mean body surface area of 1.35±0.11m2. Preoperatively, 8 patients were in New York Heart Association class II and 1 was in class III. The reason for surgery was aortic stenosis in 8 patients and aortic regurgitation due to infective endocarditis in 1 patient. Four patients had chronic renal failure and were on hemodialysis, while 1 patient had Crohn's disease. Concomitant coronary artery bypass grafting was performed in 3 patients, and tricuspid valve annuloplasty was done in 1 patient. The follow-up period was 12.0±7.71 months. No deaths occurred, but there was 1 cerebral infarction. Postoperatively, the peak pressure gradient decreased from 81.3±32.7 to 40.3±16.3mmHg (p<0.01). The mean pressure gradient also decreased significantly from 48.8±11.6mmHg to 23.9±9.32mmHg (p<0.01). Left ventricular end-diastolic diameter was 47.9±3.82mm preoperatively and 45.1±7.53mm postoperatively, showing no significant change. The left ventricular mass index also improved from 217.3±46.9 to 160±54.9g/m2 (p<0.05). The ejection fraction was 72.0±8.93% preoperatively and 67.6±6.37% postoperatively, showing no difference. Although the postoperative indexed effective orifice area (EOAI) was 0.90±0.11cm2/m2, mild patient-prosthesis mismatch (EOAI 0.77cm2/m2) was noted in 1 patient. In conclusion, the early clinical and hemodynamic performance of the 19-mm MMV in small elderly patients was acceptable.