A Digital System for Bioimpedance and Electrical Impedance Tomography Measurement System.
- Author:
Xiaoyan CHEN
;
Nana GAO
;
Huafang HUANG
- Publication Type:Journal Article
- MeSH:
Animals;
Electric Impedance;
Electrodes;
Tomography;
methods
- From:
Journal of Biomedical Engineering
2015;32(3):575-580
- CountryChina
- Language:Chinese
-
Abstract:
A digital system for bioimpedance and electrical impedance tomography (EIT) measurement controlled by an ATmega16 microcontroller was constructed in our laboratory. There are eight digital electrodes using AD5933 to measure the impedance of the targets, and the data is transmitted to the computer wirelessly through nRF24L01. The structure of the system, circuit design, system testing, vitro measurements of animals' tissues and electrical impedance tomography are introduced specifically in this paper. The experimental results showed that the system relative error was 0.42%, and the signal noise ratio was 76.3 dB. The system not only can be used to measure the impedance by any two electrodes within the frequency of 1-100 kHz in a sweep scanning, but also can reconstruct the images of EIT. The animal experiments showed that the data was valid and plots were fitting with Cole-Cole theory. The testing verified the feasibility and effectiveness of the system. The images reconstructed of a salt-water tank are satisfactory and match with the actual distribution of the tank. The system improves the effectiveness of the front-end measuring signal and the stability of the system greatly.