Retrieval of a broken guide wire transfixing the hip through the proximal femoral nail hole.
- Author:
Saurabh JAIN
1
;
Abhishek PATHAK
;
Rajeev Kant PANDEY
Author Information
1. Department of Orthopaedics, Gandhi Medical College, Bhopal, India.
- Publication Type:Case Reports
- MeSH:
Adult;
Bone Wires;
adverse effects;
Fracture Fixation, Internal;
instrumentation;
Hip Fractures;
surgery;
Humans;
Male;
Prosthesis Failure
- From:
Chinese Journal of Traumatology
2014;17(1):50-53
- CountryChina
- Language:English
-
Abstract:
Retrieval of a broken guide wire transfixing the acetabulum or with intrapelvic migration is challenging and frustrating for surgeons. We here present a case report on a method to remove a broken guide wire transfixing the acetabulum through the proximal hole of recon nail using a grasping forceps. This method is little invasive, easy, time-saving and without need for changing the inital fixation.