A real-time flow velocity estimation applied to ultrasound color display.
- Author:
Jianqi DENG
1
;
Dongquan LIU
;
Jiliu ZHOU
Author Information
1. College of Electronic Information Engineering, Sichuan University, Chengdu 610064, China.
- Publication Type:Journal Article
- MeSH:
Algorithms;
Blood Flow Velocity;
Humans;
Pulsatile Flow;
physiology;
Rheology;
instrumentation;
Ultrasonography, Doppler, Color;
instrumentation;
methods
- From:
Journal of Biomedical Engineering
2007;24(3):530-533
- CountryChina
- Language:Chinese
-
Abstract:
In this paper the theory of autocorrelation algorithm for color flow mapping is analyzed and a new way for ultrasound color flow velocity estimation and real-time display is proposed. The method sets up a mapping table which directly relates to the dynamic display range and has only 256 entries using an inverse mapping method instead of calculating the arctangent value directly. This method is ideal for software implementation and offers an interactive way to the user for changing the dynamic range of flow velocity and thus to increasing the display resolution.