The leakage radiation testing method for mammographic X-ray equipments.
- Author:
Lian-Ji XIA
1
;
Jian-Jun WANG
;
B DUPONT
;
Jane CHEN
Author Information
1. State Quality Supervision and Inspection Center of Medical X-Ray Equipments, Shenyang 110026. lianjixia@163.com
- Publication Type:Journal Article
- MeSH:
Equipment Design;
Equipment Failure Analysis;
Female;
Humans;
Mammography;
instrumentation;
methods;
Quality Control;
Radiation Dosage;
Radiation Protection;
instrumentation;
methods;
Radiometry;
instrumentation;
methods
- From:
Chinese Journal of Medical Instrumentation
2006;30(2):129-130
- CountryChina
- Language:Chinese
-
Abstract:
This article introduces a new testing method for mammographic x-ray equipments. The films placed around the x-ray tube assembly are exposured to find the location of leakage radiation and then the accurate testing for the leakage radiation of the mammographic x-ray equipment is carried out with a radiation dosimeter.