Comparison of lung function measurement values with two different apparatus.
- Author:
Feng-ying HAO
1
;
Bing-xiang YU
;
Li-xin XIE
;
Lu CAO
;
Xiang-qun FANG
;
Jian-xin WANG
;
You-ning LIU
Author Information
1. Department of Respiratory Medicine, PLA General Hospital, Beijing.
- Publication Type:Journal Article
- MeSH:
Adult;
Aged;
Female;
Humans;
Male;
Middle Aged;
Respiratory Function Tests;
instrumentation;
methods
- From:
Chinese Journal of Medical Instrumentation
2006;30(6):459-440
- CountryChina
- Language:Chinese
-
Abstract:
This paper introduces a random measurement analysis of, lung function measurement values with two different apparatus. in 41 patients. It shows that the differences are not statistically significant (P>0.05) between two apparatus measurement values except DLCO, FEF25, FEF75 in the group of normal ventilation, FVC in the group of abnormal ventilation. The two groups are both correlated closely (r> 0.9) except MMF(r=0.7725, RV r=0.808) in the normal group of ventilation, and FEF75 (r=0.58) in the abnormal group of ventilation (p<0.001). The two apparatus with different measuring theories have a good correlation.