Dosimetric comparison between helical tomotherapy and intensity-modulated radiation therapy plans for non-small cell lung cancer.
- Author:
Ling-Ling MENG
1
;
Lin-Chun FENG
;
Yun-Lai WANG
;
Xiang-Kun DAI
;
Chuan-Bin XIE
Author Information
- Publication Type:Journal Article
- MeSH: Adult; Aged; Aged, 80 and over; Carcinoma, Non-Small-Cell Lung; diagnostic imaging; radiotherapy; Humans; Male; Middle Aged; Radiography; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; methods; Radiotherapy, Intensity-Modulated; methods; Treatment Outcome
- From: Chinese Medical Journal 2011;124(11):1667-1671
- CountryChina
- Language:English
-
Abstract:
BACKGROUNDHelical tomotherapy (HT) is a new image-guided intensity-modulated radiation therapy (IMRT) technique. It is reported that HT plan for non-small-cell lung cancer (NSCLC) can give better dose uniformity, dose gradients, and protection for the lung than IMRT plan. We compared the dosimetric characteristics of HT for NSCLC with those of conventional IMRT to observe the superiority of HT.
METHODSThere was a comparative case series comprising 10 patients with NSCLC. Computed tomographic (CT) images of delineated targets were transferred to the PrecisePlan planning system (IMRT) and Tomo planning system (HT). The prescription doses were 70 Gy/33F for the gross tumor volume (GTV) and the visible lymph nodes (GTVnd), and 60 Gy/33F for the clinical target volume (CTV) and the clinical target volume of the visible lymph nodes (CTVnd). The dose restrictions for organs at risk were as follows: the maximum dose to spinal cord ≤ 45 Gy, V20 to the total lungs < 30%, V50 to the heart < 50%, and V55 to the esophagus < 50%. Both plans were evaluated by means of the dose coverage of the targets, dose-volume histograms (DVHs), and other dosimetric indices.
RESULTSThe dose coverage, conformity, and homogeneity of the targets' volumes were found to be satisfactory in both plans, but the homogeneity of the HT plan was better than that of IMRT. The high-dose radiation volume (V20-V30) to the lung and the mean lung dose (MLD) decreased (P < 0.05), but the low-dose radiation volume (V5-V10) increased slightly in the HT plan (P > 0.05). The maximum doses to the spinal cord, heart, esophagus and trachea in the HT plan were lower than those in the IMRT plan, but the differences were not statistically significant.
CONCLUSIONSThe HT plan provids better dose uniformity, dose gradients, and protection for the organs at risk. It can reduce the high-dose radiation volume for lung and the MLD, but may deliver a larger lung volume of low-dose radiation.