Prevention of restenosis in the canine coronary stents through local delivery of paclitaxel using the double-balloon perfusion catheter.
- Author:
Jin-Da WANG
1
;
Ting-Shu YANG
;
Zhi-Jun SUN
;
Jun GUO
;
Feng TIAN
;
Yun-Dai CHEN
Author Information
- Publication Type:Journal Article
- MeSH: Angioplasty, Balloon, Coronary; Animals; Catheters; Coronary Restenosis; prevention & control; Disease Models, Animal; Dogs; Injections; Paclitaxel; administration & dosage; therapeutic use; Stents
- From: Acta Academiae Medicinae Sinicae 2012;34(1):8-13
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo evaluate the safety and efficiency of local paclitaxel delivery using the double-balloon perfusion catheter to prevent restenosis in the canine coronary artery.
METHODSTwenty domestic canines underwent bare-mental stent implantation after balloon injure of the left coronary artery. A novel double-balloon perfusion catheter was used to deliver the drug locally in the canine coronary artery. In the treatment group (n = 15), paclitaxel (10 ml, 20 micromol/L) was delivered using the double-balloon perfusion catheter before stent implantation. In the control group (n = 5), 10 ml saline was delivered using the double-balloon perfusion catheter before stent implantation. The perfusion time in both groups was (26.45 +/- 5.18) s. Animals underwent coronary angiography and optical coherence tomography (OCT) 90 days after stent implantation and were sacrificed. Vessels were perfusion-fixed and morphometric analysis was performed using conventional techniques.
RESULTSCoronary angiography results showed restenosis rate in control group was significantly higher than that in treatment group (60% vs. 33.33%, P < 0.05). The parameters of OCT showed in treatment group and control group: the neointimal thickness was (0.19 +/- 0.08) mm and (0.38 +/- 0.03) mm, the neointimal area was (1.52 +/- 0.49) mm2 and (2.51 +/- 0.47) mm2, the lumen area was (3.50 +/- 0.66) mm2 and (2.78 +/- 0.57) mm2, the extent of stenosis was (30.13 +/- 8.56)% and (47.40 +/- 4.50)%, and all the variances above were significantly different between the two groups (P < 0.05). The histologic parameters showed in treatment group and control group: the neointimal thickness was (0.22 +/- 0.10) mm and (0.47 +/- 0.05) mm, the neointimal area was (1.85 +/- 0.78) mm2 and (3.43 +/- 0.25) mm2, the lumen area was (3.15 +/- 0.43) mm2 and (1.85 +/- 0.55) mm2, the extent of stenosis was (36.00 +/- 10.97)% and (65.40 +/- 8.23)%, and all the variances above were also significantly different between the two groups (P < 0.05). The stents of both the groups were fully endothelialized. No thrombus or aneurysm was found in stents.
CONCLUSIONLocal delivery of paclitaxel with the double-balloon perfusion catheter to prevent restenosis in coronary stents is safe and efficient.