An experimental study on rabbit liver by 3.0 Tesla magnetic resonance diffusion-weighted imaging.
- Author:
Tian-Ming DONG
1
;
Xi-Jie SUN
;
Xian XU
;
Jing DAI
;
Wei ZHENG
;
Ning-Yu AN
;
Xiao-Qing WANG
Author Information
- Publication Type:Journal Article
- MeSH: Animals; Diffusion Magnetic Resonance Imaging; methods; Liver; Male; Rabbits
- From: Acta Academiae Medicinae Sinicae 2012;34(1):51-57
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo explore the methodology as well as the features, quantificational index, and reference standard of 3.0 Tesla magnetic resonance (MR) diffusion-weighted imaging (DWI) on the normal rabbit's liver.
METHODSTwenty New Zealand white rabbits were enrolled and DWI was performed after anesthetics with multi-b values at 3.0 T MR scanner. Apparent diffusion coefficient (ADC) values as well as the difference between maximum and minimum ADC values, signal strength (SH), noise signal (SD), signal to noise ratio (SNR), and quality index (QI) were recorded and analyzed.
RESULTSWith b value increased, the ADC values decreased accordingly (P < 0.001). The difference between maximum and minimum ADC values with b = 1000 s/mm2 was the least (good stability), b = 600 s/mm2 was the second least, and b = 300 s/mm2 was greatest (bad stability). The SH decreased at the same time (P < 0.001), but the difference among DWI with b =600, 800, and 1000 s/mm2 was not statistically significant (P > 0.05). The SD decreased at the same time (P < 0.001), but the difference between DWI with b = 800 s/mm2 and b = 1000 s/mm2 was not statistically significant (P > 0.05). The SNR decreased at the same time (P < 0.001), but there were no significant differences between DWI with b = 600 s/mm2 and b = 800 s/mm2 or b = 600 s/mm2 and b = 1000 s/mm2 (P > 0.05). The SNR of DWI with b = 800 s/mm2 and b = 1000 s/mm2 was lower. The QI decreased at the same time (P < 0. 001) , but the difference between DWI with b = 800 s/mm2 and b = 1000 s/mm was not statistically significant (P > 0.05).
CONCLUSIONWhen 3.0 T MR DWI is applied for rabbit liver, it is better to use b = 600 s/mm2 for reducing scanning time and assuring better diffusion weights, quantity of images, and stability of ADC measurement.