Establishment of an isolated porcine liver machine perfusion model for high-intensity focused ultrasound studies.
- Author:
Rui BU
1
;
Jianzhong ZOU
;
Li YIN
;
Feng WU
;
Shihe MENG
;
Fang LIU
;
Xin ZHAO
;
Jiao JIAO
;
Han YANG
Author Information
- Publication Type:Journal Article
- MeSH: Animals; Equipment Design; Extracorporeal Circulation; instrumentation; methods; Hemodynamics; Liver; blood supply; diagnostic imaging; Liver Circulation; physiology; Swine; Ultrasonography
- From: Journal of Southern Medical University 2012;32(11):1564-1567
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo explore the feasibility of establishing an isolated porcine liver machine perfusion model and assess its value in high-intensity focused ultrasound studies.
METHODSTwenty-one isolated porcine livers were perfused with autologous blood for 4 h through dual vessels (portal vein and hepatic artery) cannulation using an extracorporeal circulation machine under a sub-normothermic perfusion condition. The perfusion model was assessed by monitoring the liver color, texture, liver weight gain, hemodynamic parameters, color Doppler flow imaging, bile output and histopathology.
RESULTSNineteen isolated porcine livers were successfully cannulated with dual vessels, and failure of hepatic artery intubation occurred in two porcine livers. After machine perfusion for 4 h, the isolated livers maintained a soft texture with stable hemodynamic levels within relative normal physiological ranges. The bile output was more than 3 ml/h within the initial 3 h of perfusion. Histopathological examination demonstrated no morphological or structural changes of the liver tissues.
CONCLUSIONThe isolated porcine liver perfusion model is stable and feasible, and can be used for high-intensity focused ultrasound studies.