Assessment of the value of the clinical application of special infusion pump pipe.
- Author:
Lei SHAO
1
;
Kun YANG
;
Ying YUE
;
Bo ZHENG
Author Information
1. Shanghai Tongji Hospital Equipment Materials Department, Shanghai 200065. lei_shao@163.com
- Publication Type:Journal Article
- MeSH:
Equipment Safety;
Humans;
Infusion Pumps;
utilization
- From:
Chinese Journal of Medical Instrumentation
2012;36(4):302-303
- CountryChina
- Language:Chinese
-
Abstract:
The clinical application value of infusion pumps dedicated infusion device was evaluated through the comparison on the accuracy between dedicated and non-specific infusion devices. Experiments were conducted by using five test points each repeated three times to detecting accuracy of the infusion of the three kinds of infusion devices. Through the analysis of test results, we found out that the infusion accuracy of special and non-dedicated infusion devices in the experiment, exists significant differences (P < 0.05), and we believe that the use of infusion pumps dedicated infusion device can guarantees safety and reduces medical risks in clinical.