Analysis residues of perfluorooctane used in ophthalmic operation.
- Author:
Minzhu WANG
1
;
Li ZHANG
;
Menglan QIAN
Author Information
1. Medical Device Supervising and Testing Center of Hangzhou, State Food and Drug Administration, Hangzhou 310009. wangminzhu@21cn.com
- Publication Type:Journal Article
- MeSH:
Fluorocarbons;
analysis;
chemistry;
Ophthalmology;
instrumentation;
methods;
Quality Control
- From:
Chinese Journal of Medical Instrumentation
2012;36(5):373-374
- CountryChina
- Language:Chinese
-
Abstract:
Analyzed the purity of the Perfluorooctane used for Ophthalmic operation using the FID (Flame Ionization Detector) method after the qualitative analysis of that. Research and discussion of the 1--hydrogen perfluorocarbon octane, which is the typical impurity of Perfluorooctane. The standard recovery of the 1--hydrogen perfluorocarbon octane is 93%-97% within the range of linearity, and the RSD (relative standard deviation) is below 2%. The above test method has the advantages of easy operation, stability and accuracy. So it is a good reference support to bring the specification of quality control of the Perfluorooctane which is used for Ophthalmic operation into YY of medical device.