Clinical implementation of dose reconstruction and dose-guided intensity modulated radiotherapy for helical tomotherapy.
- Author:
Weirong YAO
1
;
Shouping XU
;
Du LEI
;
Chuanbin XIE
;
Lin MA
Author Information
- Publication Type:Journal Article
- MeSH: Humans; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; methods; Radiotherapy, Intensity-Modulated; methods; Tomography, Spiral Computed; methods
- From: Chinese Journal of Medical Instrumentation 2012;36(5):375-377
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo implement dose reconstruction and dose-guided intensity modulated radiotherapy for helical tomotherapy.
METHODSDose reconstruction was implemented on adaptive helical tomotherapy with the online megavoltage CT (MVCT) imaging from a patient with nasopharyngeal cancer. The differences of isodose line between actual and planned deposition were analysis in 3D distribution, on which the hot spot and cold spot were lined. The dose delivered to these areas was modulated in later fractions to keep the planned requirement.
RESULTSThe differences between actual and planned isodose line were shown on the image visually. The modulation to the hot spot and cold spot in later fraction corrected the incorrectly delivered dose to achieve the requirement of primary plan.
CONCLUSIONSThe dose reconstruction and dose-guided intensity modulated radiotherapy can be implemented in adaptive helical tomotherapy.