Design and control of a simulating cardiovascular system.
- Author:
Yi LIU
1
;
Ming YANG
;
Shiyang LI
;
Zhejun ZHENG
Author Information
1. Department of Instruments Science and Engineering, Shanghai Jiaotong University, Shanghai 200240, China. richieliu@sjtu.edu.cn
- Publication Type:Journal Article
- MeSH:
Cardiovascular Physiological Phenomena;
Computer Simulation;
Equipment Design;
Heart, Artificial;
Heart-Assist Devices;
Hemodynamics;
Models, Cardiovascular
- From:
Journal of Biomedical Engineering
2010;27(1):165-169
- CountryChina
- Language:Chinese
-
Abstract:
In order to study the cardiovascular hemodynamic characteristics and evaluate the artificial heart, especially the compression cardiac assist devices, we put forward a simulating device which is made of pump, valve, tubes and controller, and can be used to imitate cardiovascular system. Moreover, in this essay, we put forth an algorithm to draw pressure-volume loop with the parameters measured, including liquid pressure, volume and air pressure. The changes in the frequency and duty of control signal and in the angle of proportional valve can symbolize cardiovascular system in different heart rates, with different systole period, and at different level, respectively. The result of simulating device experiment proved to be coincident with physiology.