Evaluation of the performance of the electrochemiluminescent immunoassay system of Elecsys 2010.
- Author:
Weishi QIN
1
Author Information
1. Department of Nuclear Medicine, West China Hospital, Sichuan University, Chengdu 610041.
- Publication Type:Journal Article
- MeSH:
Electrochemistry;
instrumentation;
Evaluation Studies as Topic;
Humans;
Immunoassay;
instrumentation;
standards;
Luminescent Measurements;
Sensitivity and Specificity;
Thyrotropin;
blood
- From:
Journal of Biomedical Engineering
2002;19(4):673-675
- CountryChina
- Language:Chinese
-
Abstract:
Performance of the electrochemiluminescent immunoassay system of Elecsys 2010 was evaluated by quantitative assay. The functional sensitivity of the TSH assay was 0.004 microU/ml. The extra-assay imprecision was RSD = 1.45%-4.65% in the range of TSH concentrations of 0.43-88.15 microU/ml. The intra-assay imprecision was RSD = 1.38%-3.14% in the range of TSH concentrations of 0.41-87.28 microU/ml. The recovery of TSH was in the range of 98.6%-104.4%. The correlation between theoretical values and obtained values of TSH was adequate(r = 0.999). We conclude that the Elecsys 2010 is provably satisfactory in sensitivity, precision, accuracy, reliability, practicability and rapidity.