Clinical application of robotic tele-manipulation system in stereotactic surgery.
- Author:
Zeng-Min TIAN
1
;
Wang-Sheng LU
;
Tian-Miao WANG
;
Du LIU
;
Yan CHEN
;
Guo-Lai ZHANG
;
Quan-Jun ZHAO
;
Mang-Mang BAI
;
Feng YIN
Author Information
- Publication Type:Journal Article
- MeSH: Adolescent; Adult; Aged; Brain; pathology; surgery; Brain Diseases; surgery; Child; Child, Preschool; Female; Follow-Up Studies; Humans; Male; Middle Aged; Reproducibility of Results; Retrospective Studies; Robotics; methods; Stereotaxic Techniques; Surgery, Computer-Assisted; Treatment Outcome
- From: Chinese Journal of Surgery 2007;45(24):1679-1681
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo assess the clinical usefulness, accuracy, and safety of tele-manipulation for frameless stereotactic surgery using the CAS-R-5 robot system.
METHODSWe prospectively evaluated 32 patients underwent tele-manipulation of frameless stereotactic operations from Sep. 2005 to Sep. 2006. Tele-manipulations were performed via a digital data network by a neurosurgeon in Beijing while the patients were located in Yan'an. The distance is 1300 kilometers away. The accuracy of location and improvement of symptom were observed after operation. The period of follow-up was from 3 to 14 months (the average was 12 months).
RESULTSThe surgical operations in 32 cases were successful. Remote fiducial registration was performed with a mean accuracy of 1. 50 mm and the standard difference were 0.32 mm between the planned and actual target. There were no complications.
CONCLUSIONSDiagnosis and treatment for intracranial disease by tele-manipulation frameless stereotactic surgeries are reliable and safe.