Abstract:
OBJECTIVETo observe the efficacy and adverse reactions of autologous PBSC collection when the autoPBSC procedure and MNC procedure of COBE Spectra cell separator and the MNC procedure of Spectra Optia cell separator were used.
METHODSThe autologous perepheral blood hematopoietic stem cells from 41 patients were collected by using autoPBSC procedure and MNC procedure of COBE Spectra blood cell separator and MNC procedure of Spectra Optia blood cell separator. The numbers of MNC and CD34cells collected by 3 collected procedure, the difference of hemoglobin (Hb) drop and platelet decrease, and the adverse reaction of patients were observed.
RESULTSWhen the whole blood processing and the collection time were basically same among these 3 groups, the MNC counts collected by MNC procedure of COBE Spectra and Spectra Optia were higher than that of AutoPBSC procedure of COBE Spctra, but the CD34cell count was lower than that collected by AutoPBSC procedure (P< 0.05). The final product volume collected by MNC procedure of COBE Spectra and Spectra Optia was bigger than that collected by AutoPBSC procedure. In comprission with MNC procedure of COBE Spectra cell seperator, the CD34count collected by MNC procedure of Spectra Optia Seperator did not show significant difference, but the CD34cell count collected by MNC procedure of Spectra Optia was higher than that collected by MNC procedure of COBE Spectra cell separator (P<0.05). The platelet count and hemoglobin level collected by MNC procedure of Spectra Optia were lower than those before collection. The adverse reactions in the 3 procedures were similar, and the patients could tolerate them.
CONCLUSIONThe AutoPBSC procedure of COBE Spectra and MNC procedure of Spectra Optia are better than MNC procedure of COBE Spectra for autologous peripheral blood hematopoietic stem cells collection. The loss of blood platelet and hemoglobin after collection is lowest in MNC procedure of Spectra Optia.