Comparative test on puncture coring of two different needles used with the implantable drug-supplying device.
- Author:
Min WAN
1
;
Ping WU
;
Jinzi SONG
;
Xin YU
;
Pengtao MOU
Author Information
1. Ji' nan Quality Supervision and Inspection Center for Medical Devices affiliated to SFDA, Ji' nan, Shandong, 250101. annvickey@126.com
- Publication Type:Journal Article
- MeSH:
Infusion Pumps, Implantable;
classification;
Needles
- From:
Chinese Journal of Medical Instrumentation
2010;34(6):445-447
- CountryChina
- Language:Chinese
-
Abstract:
A comparison test on the number of puncture coring is conducted using the normal needle and the Huber needle in the injection area of the implantable drug-supplying device separately. The result indicates that the number of coring using the Huber needle is much less than that using the normal needle during the puncturing. So it is suggested to popularize the Huber needle in the drug transfusion of the implantable drug-supplying device.