Comparison of two coils for MR angiography based on the SDNR method.
- Author:
Longchen WANG
1
;
Bin LI
;
Yunfeng XIAO
;
Ruihua QIAO
Author Information
1. BME Department, The Sixth People's Hospital Affiliated to Shanghai Jiao Tong University, Shanghai, 200233. wanglch09@gmail.com
- Publication Type:Journal Article
- MeSH:
Adult;
Cerebral Angiography;
methods;
Humans;
Magnetic Resonance Angiography;
methods;
Male;
Signal-To-Noise Ratio
- From:
Chinese Journal of Medical Instrumentation
2011;35(4):256-259
- CountryChina
- Language:Chinese
-
Abstract:
In this paper, signal-to-noise ratio (SNR) and signal different-to-noise ratio (SDNR) methods were used to compare image quality using two different radiofrequency coils. The two coil types included an eight-element phased-array coil and a quadrature birdcage head coil with endcap and the comparison studies performed on a uniform cylindrical phantom and volunteer respectively. The results showed phased-array coil have advantages and proved effectiveness of the method in the evaluation and selection of coils.