Calculation of inspiratory tidal volume of Datex Ohmeda 7100 anesthetic machine.
- Author:
Wen ZHOU
1
Author Information
1. The 97th Hospital of PLA, Xuzhou, 221004. xzzhouwen@yahoo.cn
- Publication Type:Journal Article
- MeSH:
Anesthesia;
methods;
Monitoring, Intraoperative;
instrumentation;
methods;
Surgical Equipment;
Tidal Volume
- From:
Chinese Journal of Medical Instrumentation
2009;33(3):223-224
- CountryChina
- Language:Chinese
-
Abstract:
According to the basic work principle of Datex Ohmeda 7100 anesthetic machine, the calculation formula of inspiratory tidal volume is deduced in the paper. The formula is verified by clinical application. The summary how to direct the applying and maintaining anesthetic machine and its monitoring instrument by using formula is presented.