Micro-CT system based on the flat panel detector for small animal imaging.
- Author:
Xiao-Quan YANG
1
;
Yong DENG
;
Zi-Lin DENG
;
Qing-Ming LUO
;
Hui GONG
Author Information
1. Wuhan National Laboratory for Optoelectronics, Huazhong University of Science and Technology.
- Publication Type:Journal Article
- MeSH:
Animals;
Equipment Design;
Rabbits;
X-Ray Microtomography;
instrumentation
- From:
Chinese Journal of Medical Instrumentation
2009;33(4):255-258
- CountryChina
- Language:Chinese
-
Abstract:
A high resolution Micro-CT system for small animal imaging is introduced in this paper. Micro-focus X-ray tube with focal diameter of 100 microm and flat panel detector with imaging area of 13cm x 13cm are adopted in this system. The data acquired in rotation scanning is reconstructed with cone beam algorithm. The resolving power of the detector is measured to be 31 lp/cm at 10% of the MTF. The resolution of the Micro-CT system could achieve 185 m when the magnification factor is 1.94. Thighbone of a rabbit is used as sample imaging with the system. The trabecular bone could be imaged clearly. And the ability of small animal imaging of the system has been demonstrated.