Determination of polysorbate 80 in Reduning injection by HGPC-ELSD.
- Author:
Juan SHEN
;
Qiao ZHANG
;
Jia-Chun LI
;
Yu-An BI
;
Zhen-Zhong WANG
;
Wei XIAO
- Publication Type:Journal Article
- MeSH: Calibration; Chromatography, High Pressure Liquid; methods; Drugs, Chinese Herbal; chemistry; Injections; Polysorbates; analysis; Reproducibility of Results; Time Factors
- From: China Journal of Chinese Materia Medica 2014;39(15):2915-2917
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish the method for determining polysorbate 80 in Reduning injection by HPLC-ELSD, and to control the mass of polysorbate 80 in Reduning injection.
METHODIt was performed by HGPC-ELSD with TOSHTSK-GEL G4000PWxl (7.8 mm x 300 mm, 10 μm). Water was used as mobile phase, the flow rate was 0.7 mL x min(-1), and the temperature was set at 30°C. The evaporated light scattering detector was adopted. The drift tube temperature was 55°C, and nitrogen was used as carrier gas, with the flow rate of 2.0 L x min(-1) and gain of 1.0.
RESULTThe calibration curve showed good linearity of polysorbate 80 in the test range from 1.01 to 15.20 g x L(-1) (r2 = 0.999 3). The recovery rate was 98.10% with RSD of 2.0%.
CONCLUSIONThe method is simple, rapid, accurate and reliable and suitable for the determination of polysorbate 80 in Reduning injection.